Stratos LV/LV-T Technical Manual 73
Figure 3. Scan Hysteresis
Scan hysteresis has been incorporated to promote intrinsic
cardiac rhythm and may reduce device energy consumption.
The number of scan interval is programmable, OFF, 1…(1)…15
cycles.
NOTES:
Scan Hysteresis is not active during the programmed Night
Mode.
Scan Hysteresis is only available when Hysteresis is
selected on.
Magnet application (closing of reed switch) suspends 180
consecutive event counter independent of the magnet effect.