Stratos LV/LV-T Technical Manual 141
Lead Compatibility
The Stratos CRT-Ps have been designed for connection with a
bipolar lead in the atrium and two bipolar leads in the ventricle.
All connections are IS-1 compatible. Appropriate adapters (e.g.,
A1-A) should be fitted when using leads with a different
connection.
NOTE:
Connecting systems with a 3.2 mm configuration that do not
expressly claim to agree with the IS-1 dimensions generally
have to be regarded as incompatible with IS-1 connectors
and can only be used with BIOTRONIK products together
with an appropriate adapter. For questions regarding lead-
device compatibility, consult your BIOTRONIK
representative.
In case of device replacement, ensure that the existing lead
connector and leads are not damaged.
CAUTION
Lead / CRT-P Compatibility – Because of the numerous
available 3.2-mm configurations (e.g., the IS-1 and VS-1
standards), lead/ CRT-P compatibility should be confirmed
with the CRT-P and/or lead manufacturer prior to the
implantation of the system.
IS-1, wherever stated in this manual, refers to the international
standard, whereby leads and generators from different
manufacturers are assured a basic fit. [Reference ISO 5841-
3:1992(E)].